2019/08/07

  If you utilize NIR spectroscopy for research, process or diagnostics, NIRQuest is a less costly, less complex alternative to FT-IR and comparable technologies.

  NIRQuest delivers a high-performance optical bench with low-noise electronics and more grating options than we've ever offered.

  This compact, multi-interface spectrometer is available in version for 900-1700 nm, 900-2050 nm, 900-2200 nm and 900-2500 nm.

  NIRQuest's modular design delivers more customization for a wider variety of applications including medical diagnostics, pharmaceuticals analysis, environmental monitoring and process control.

  NIRQuest for Laser Analysis

  NIRQuest is versatile enough to be configured for high-resolution laser and optical fiber characterization. The NIRQuest delivers reliable optical resolution to <0.25 nm (FWHM) and is easily customized for your high-resolution spectral applications.

  Thulium Fiber Laser

  Courtesy of Townes Laser Institute at the University of Central Floridas College of Optics & Photonics

  Turn NIRQuest into a Network Spectrometer

  When purchased with the Remora Network Adapter users can communicate with the NIRQuest over Ethernet or Wi-Fi. Use with SpectraSuite Software, and turn your device into a network accessible spectrometer. Remotely capture spectral data from anywhere Plus, Remora allows multiple users to access data and experiment settings without the need for additional equipment. Remora's web-interface works with all popular web browsers and allows you to monitor data from a PDA, iPhone, and most internet ready devices. - Learn more

At a Glance

NIRQuest512NIRQuest512-2.2NIRQuest256-2.1NIRQuest256-2.5
Detector:Hamamatsu G9204-512 InGaAs linear arrayHamamatsu G9204-512W InGaAs linear arrayHamamatsu G9206-256 InGaAs linear arrayHamamatsu G9208-256 InGaAs linear array
Detector Range:850-1700 nm900-2200 nm900-2100 nm900-2550 nm
Useable range1:900-1700 nm900-2200 nm900-2050 nm900-2500 nm
Pixels:512512256256
Pixel size:25 m x 500 m25 m x 250 m50 m x 500 m50 m x 500 m
Entrance aperture:
25 m wide slit (standard); also, custom options (10, 50, 100 or 200 m wide slits) or fiber (no slit)
25 m wide slit (standard); also, custom options (10, 50, 100 or 200 m wide slits) or fiber (no slit)
25 m wide slit (standard); also, custom options (50, 100 or 200 m wide slits) or fiber (no slit)25 m wide slit (standard); also, custom options (50, 100 or 200 m wide slits) or fiber (no slit)
Fiber optic connector:SMA 905SMA 905SMA 905SMA 905
Grating options (standard):Grating NIR3, 150 l/mm, 900-1700 nmGrating NIR2, 100 l/mm, 900-2050 nmGrating NIR2, 100 l/mm, 900-2050 nmGrating NIR1, 150 l/mm, 1075-2500 nm
Grating options (custom):Gratings NIR10, NIR11, NIR12, NIR13 and NIR14Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13
Longpass filter2:OF1-RG830 longpass NIR filter (optional)OF1-RG830 longpass NIR filter (optional)OF1-RG830 longpass NIR filter (optional)OF1-RG830 longpass NIR filter (optional)
Wavelength range:900-1700 nm 
w/Grating NIR3
900-2150 nm 
w/Grating NIR2
900-2050 nm 
w/Grating NIR2
900-2500 nm 
w/Grating NIR1
Optical resolution (FWHM)3:~3.1 nm w/25 m slit~4.6 nm w/25 m slit~7.6 nm w/25 m slit~ 9.5 nm w/25 m slit
Signal-to-noise ratio at full signal4:>15000:1 @ 100 ms integration
>13000:1 @ 1000 ms integration
>10000:1 @ 100 ms integration10000:1 @ 100 ms integration7500:1 @ 10 ms integration
A/D resolution:16-bit16-bit16-bit16-bit
Dark noise:6 RMS counts @ 100 ms
12 RMS counts @ 
1000 ms
6 RMS counts @ 100 ms
12 RMS counts @ 
250 ms
6 RMS counts @ 100 ms
12 RMS counts @ 250 ms
8 RMS counts @ 10 ms
12 counts RMS @ 30 ms
Dynamic range:15 x 10^6 (system); 5000:1 for a single acquisition15 x 10^6 (system); 1000:1 for a single acquisition5 x 10^6 (system); 5000:1 for a single acquisition3  x 10^5  (system); 5000:1 for a single acquisition
Integration time5:1 ms-10 seconds1 ms-2000 ms1 ms-2000 ms1 ms-60 ms
Corrected linearity:>99.8%>99.8%>99.8%>99.8%
Computer interfaces:
USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud
Wireless (Wi-Fi) interface:
Yes, with Remora adapter
Wi-Fi range:
25 meters in free space

1 Useable range?is defined in the context of the NIRQuest models detector response and its typical grating response. For example, the 512-element detector has response at 850 nm, but grating response begins at 900 nm. The G9206 256-element detector response is sensitive to TEC temperature, and has response only to 2050 nm when the TEC is set to -20 C. The G9208 256-element detector has response to 2550 nm, but the grating efficiency drops off at 2500 nm.
2 Other filter options are available for order-sorting in the NIRQuest256-2.1 and NIRQuest256-2.5. Please contact an Applications Scientist for details.
3 Optical resolution (FWHM) depends on grating and slit selection.
4 SNR will decrease at longer integration times.

5 Maximum integration times are defined as the longest amount of time one can integrate the spectrometer before the dark level rises to half of full scale.
 